胡越, 孟祥娟, 张晓梅, 刘晓东. 基于测量不确定度表示指南法和蒙特卡洛法评价血铅测量的不确定度J. 职业卫生与应急救援, 2026, 44(4): 535-539,545. DOI: 10.16369/j.oher.issn.1007-1326.2026.250531
引用本文: 胡越, 孟祥娟, 张晓梅, 刘晓东. 基于测量不确定度表示指南法和蒙特卡洛法评价血铅测量的不确定度J. 职业卫生与应急救援, 2026, 44(4): 535-539,545. DOI: 10.16369/j.oher.issn.1007-1326.2026.250531
HU Yue, MENG Xiangjuan, ZHANG Xiaomei, LIU Xiaodong. Evaluation of the uncertainty of blood lead determination using GUM and Monte Carlo methodsJ. Occupational Health and Emergency Rescue, 2026, 44(4): 535-539,545. DOI: 10.16369/j.oher.issn.1007-1326.2026.250531
Citation: HU Yue, MENG Xiangjuan, ZHANG Xiaomei, LIU Xiaodong. Evaluation of the uncertainty of blood lead determination using GUM and Monte Carlo methodsJ. Occupational Health and Emergency Rescue, 2026, 44(4): 535-539,545. DOI: 10.16369/j.oher.issn.1007-1326.2026.250531

基于测量不确定度表示指南法和蒙特卡洛法评价血铅测量的不确定度

Evaluation of the uncertainty of blood lead determination using GUM and Monte Carlo methods

  • 摘要:

    目的 血铅检测属于痕量分析,为提高其测定结果的准确性和可靠性,开展测量不确定度研究,探讨适用于血铅测定的不确定度评定方法。

    方法 依据GBZ/T 316.2—2018《血中铅的测定第2部分:电感耦合等离子体质谱法》,分别采用测量不确定度表示指南法(GUM)和蒙特卡洛法(MCM)对冻干全血有证标准物质(295 ± 30)μg/L中铅含量的测定结果进行不确定度评估,并进行对比分析。

    结果 GUM得到的结果为(297±12)μg/L;MCM得到的结果为(297 ± 11)μg/L,两者计算结果相近。

    结论 当测量模型为线性时,GUM和MCM评价结果高度一致,但GUM易发生不确定度分量漏估或重复计算的问题,MCM可有效避免这种误差,提高痕量血铅检测中测量不确定度的准确性和可靠性。

     

    Abstract:

    Objective Blood lead detection falls under trace analysis. To improve the accuracy and reliability of blood lead determination, the study on measurement uncertainty was conducted to explore suitable evaluation methods for blood lead determination.

    Methods Following the standard GBZ/T 316.2—2018 Determination of Lead in Blood—Part 2: Inductively Coupled Plasma Mass Spectrometry, the measurement uncertainty associated with the determination of lead content in a certified reference material of lyophilized whole blood (295 ± 30) μg/L was evaluated using the guide to the expression of uncertainty in measurement (GUM) and the Monte Carlo method (MCM), and a comparative analysis was conducted.

    Results The result obtained by the GUM was (297 ± 12) μg/L, while that by MCM was (297 ± 11) μg/L. The two methods gave closely comparable outcomes.

    Conclusions The GUM and MCM evaluations were highly consistent when the measurement model was linear. However, the GUM was prone to under- or double-estimation of uncertainty components, whereas MCM can effectively avoid such errors and improve the accuracy and reliability of measurement uncertainty in trace blood lead detection.

     

/