唐斯, 叶立和, 蔡小璇, 张凤华, 钟苑琴, 李思燕. 复合基体改进剂石墨炉原子吸收法直接测定尿铊J. 职业卫生与应急救援, 2020, 38(5): 531-535. DOI: 10.16369/j.oher.issn.1007-1326.2020.05.026
引用本文: 唐斯, 叶立和, 蔡小璇, 张凤华, 钟苑琴, 李思燕. 复合基体改进剂石墨炉原子吸收法直接测定尿铊J. 职业卫生与应急救援, 2020, 38(5): 531-535. DOI: 10.16369/j.oher.issn.1007-1326.2020.05.026
TANG Si, YE Lihe, CAI Xiaoxuan, ZHANG Fenghua, ZHONG Yuanqin, LI Siyan. Direct determination of thallium in urine by GFAAS with compound matrix modifierJ. Occupational Health and Emergency Rescue, 2020, 38(5): 531-535. DOI: 10.16369/j.oher.issn.1007-1326.2020.05.026
Citation: TANG Si, YE Lihe, CAI Xiaoxuan, ZHANG Fenghua, ZHONG Yuanqin, LI Siyan. Direct determination of thallium in urine by GFAAS with compound matrix modifierJ. Occupational Health and Emergency Rescue, 2020, 38(5): 531-535. DOI: 10.16369/j.oher.issn.1007-1326.2020.05.026

复合基体改进剂石墨炉原子吸收法直接测定尿铊

Direct determination of thallium in urine by GFAAS with compound matrix modifier

  • 摘要:
    目的 建立以复合基体改进剂测定尿铊的改良石墨炉原子吸收光谱法。
    方法 以质量浓度0.700 g/L氯化钯、6.00%(体积比)曲拉通X-100、0.60%(体积比)硝酸作为复合基体改进剂,对尿样进行1:1稀释,应用石墨炉原子吸收光谱法直接测定尿铊。
    结果 尿铊在0.31~100.00 μg/L定量测定范围内与吸光度呈良好的线性关系,检出限为0.31 μg/L,定量下限为1.02 μg/L,以尿样稀释2倍计,尿铊的最低检出浓度为0.62 μg/L,定量浓度为2.04 μg/L。加标回收率为99.33%~107.00%,批内、批间精密度分别为0.99%~2.09%和1.36%~3.39%。尿样在4℃下可至少保存7 d。
    结论 该方法操作简单、分析快速、灵敏度高、精密性和准确性好,适用于职业性接触人群以及急慢性铊中毒者尿中铊含量的测定。

     

/

返回文章
返回