王彦梅, 蒋元强, 王桂敏, 江松, 张路钱. 上海某区一CT机房周围剂量当量率异常情况调查分析[J]. 职业卫生与应急救援, 2020, 38(5): 490-492. DOI: 10.16369/j.oher.issn.1007-1326.2020.05.013
引用本文: 王彦梅, 蒋元强, 王桂敏, 江松, 张路钱. 上海某区一CT机房周围剂量当量率异常情况调查分析[J]. 职业卫生与应急救援, 2020, 38(5): 490-492. DOI: 10.16369/j.oher.issn.1007-1326.2020.05.013
WANG Yanmei, JIANG Yuanqiang, WANG Guimin, JIANG Song, ZHANG Luqian. Investigation on abnormal dose equivalent rate around a CT room in a district of Shanghai[J]. Occupational Health and Emergency Rescue, 2020, 38(5): 490-492. DOI: 10.16369/j.oher.issn.1007-1326.2020.05.013
Citation: WANG Yanmei, JIANG Yuanqiang, WANG Guimin, JIANG Song, ZHANG Luqian. Investigation on abnormal dose equivalent rate around a CT room in a district of Shanghai[J]. Occupational Health and Emergency Rescue, 2020, 38(5): 490-492. DOI: 10.16369/j.oher.issn.1007-1326.2020.05.013

上海某区一CT机房周围剂量当量率异常情况调查分析

Investigation on abnormal dose equivalent rate around a CT room in a district of Shanghai

  • 摘要:
    目的 以实际工作中存在的机房辐射防护问题为借鉴,为医疗卫生机构科学实施放射诊疗防护管理提供依据。
    方法 通过现场监测,发现CT机房的机房防护门周围剂量当量率出现异常,根据勘查情况提出整改建议,并比较整改前后的周围剂量当量率变化。
    结果 因年久失修造成机房防护移门结构出现松动现象,周围剂量当量率为4.631~6.750 μSv/h,机房防护门外人员可能受到照射的年有效剂量值可达0.251~0.366 mSv(周有效剂量为4.827~7.036 μSv)。维修整改完成后,机房防护门周围剂量当量率为0.237~0.275 μSv/h,机房防护门外人员可能受到照射的年有效剂量值下降为0.013~0.015 mSv(周有效剂量为0.247~0.287 μSv),符合国家标准限值要求。
    结论 医疗机构应定期对机房防护门进行安全检查和维修,建立健全放射防护管理工作的长效机制。

     

/

返回文章
返回